A-6
A.2.8 Signal Output
Auxiliary Output
Standard
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
ECG Analog Output
Bandwidth (-3dB; reference
frequency:10Hz)
Diagnostic mode:
Monitor mode:
Surgical mode:
0.05 to 150 Hz
0.5 to 40 Hz
1 to 20 Hz
Max transmission delay 25 ms (in diagnostic mode, and with Notch off)
Sensitivity 1 V/mV ±5%
PACE rejection/enhancement No pace rejection or enhancement
IBP Analog Output
Bandwidth (-3dB; reference
frequency:1Hz)
8 Hz, 12.5 Hz and 20 Hz are optional.
Max transmission delay 60 ms (with Notch off)
Sensitivity 1 V/100 mmHg ±5%
Nurse Call Signal
Output mode Relay
Electrical requirements 60W, 2A, 36VDC, 25VAC
Conducting impedance <1
Isolation voltage 1500 VAC
Contact type Normally open or normally contact (optional)
Defib Sync Pulse
Max time delay 35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, providing a maximum of 1 mA output
current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width 100 ms ±10%
Output impedance 50
Rising and falling time <3 ms
VGA
Video signals
RGB: 0.7 Vp-p/75;
Horizontal/vertical synchronization: TTL level