A-15
PR from Nellcor SpO
2
Module
Measurement range 20 to 300 bpm
Resolution 1 bpm
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm, not specified
Refreshing rate 1 s
SPO2 averaging time 8 s, 16 s
PR from IBP Module
Measurement range 25 to 350 bpm
Resolution 1 bpm
Accuracy ±1 bpm or ±1%, whichever is greater
Refreshing rate 1 s