A-16
A.5.5 NIBP
Standards
Meet standards of EN60601-2-30/IEC60601-2-30, EN1060-1,
EN1060-3, EN1060-4 and SP10
Technique Oscillometry
Mode of operation Manual, Auto and STAT
Auto mode repetition intervals 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240 or 480 min
Typical measurement time 20 to 40 s (depending on the heart rate and motion artifact)
STAT mode cycle time 5 min
Adult, pediatric: 180 s
Measurement time protection
Neonate: 90 s
Adult Pediatric Neonate
Systolic: 40 to 270 40 to 200 40 to 135
Diastolic: 10 to 210 10 to 150 10 to 100
Measurement ranges
(mmHg)
Mean: 20 to 230 20 to 165 20 to 110
Accuracy
Max mean error: ±5 mmHg
Max standard deviation: 8 mmHg
Resolution 1mmHg
Defaults Range Step
Adult 160 mmHg 80 to 280
Pediatric 140 mmHg 80 to 210
Cuff inflation pressure
Neonate 90 mmHg 60 to 140
10 mmHg
Software overpressure
protection
Adult:
Pediatric:
Neonate:
297±3 mmHg
240±3 mmHg
147±3 mmHg
PR
Measurement range 40 to 240 bpm
Resolution 1 bpm
Accuracy ±2bpm or ±2%, whichever is greater
Assisting Venous Puncture
Pressure range
Adult:
Pediatric:
Neonate:
20 to 120 mmHg
20 to 80 mmHg
20 to 50 mmHg
Duration
Adult/Pediatric:
Neonate:
170 s
85 s