A-17
Alarm limit Range (mmHg) Step (mmHg)
Sys High
Adult: (low limit+5) to 270
Pediatric: (low limit+5) to 200
Neonate: (low limit+5) to 135
Sys Low 40 to (high limit-5)
Mean High
Adult: (low limit+5) to 230
Pediatric: (low limit+5) to 165
Neonate: (low limit+5) to 110
Mean Low 20 to (high limit-5)
Dia High
Adult: (low limit+5) to 210
Pediatric: (low limit+5) to 150
Neonate: (low limit+5) to 100
Dia Low 10 to (high limit-5)
5
*Measurement accuracy verification: In adult and pediatric modes, the blood pressure
measurements measured with this device are in compliance with the American National Standard
for Electronic or Automated Sphymomanometers (ANSI/AAMI SP10-1992) in terms of mean
error and stardard deviation by comparing with intra-arterial or auscultatory measurements
(depending on the configuration) in a typical patient population. For auscultatory reference, the
5th Korotkoff sound was used to determine the diastolic pressure.
In neonatal mode, the blood pressure measurements measured with this device are in compliance
with the American National Standard for Electronic or Automated Sphymomanometers
(ANSI/AAMI SP10-1992 and AAMI/ANSI SP10A-1996) in terms of mean error and stardard
deviation by comparing with intra-arterial measurements (depending on the configuration) in a
typical patient population.