A-18
A.5.6 Temp
Standards Meet standard of EN12470-4
Technique Thermal resistance
Measurement range
0 to 50 °C (32 to 122 °F)
Resolution
0.1 °C
Accuracy
±0.1 °C or ±0.2 °F (without probe)
Refreshing rate 1 s
Time for accurate
measurement
Body surface: <100 s
Body cavity: <80 s
Alarm limit Range Step
T1/T2 High
(low limit +1) to 50 °C
(low limit +1.8) to 122 °F
T1/T2 Low
0 to (high limit -1) °C
32 to (high limit -1.8) °F
TD High
0 to 50 °C
0 to 90 °F
0.1 °C
0.1 °F
A.5.7 IBP
Standards Meet standard of EN60601-2-34/IEC60601-2-34.
Technique Direct invasive measurement
IBP
Measurement range -50 to 300 mmHg
Resolution 1 mmHg
Accuracy ±2% or ±1 mmHg, whichever is greater (without sensor)
Refreshing rate 1 s
Zero adjustment
Range:
Accuracy:
Drift:
±200 mmHg
±1 mmHg
<0.15 mmHg/°C;
Less than ±1 mmHg within 24 hours
Pressure transducer
Excitement voltage 5 VDC, ±2%
Sensitivity 5 µV/V/mmHg
Impedance range 300 to 3000Ω