A-19
Alarm limit Range (mmHg) Step (mmHg)
Sys High
Mean High
Dia High
(low limit + 2) to 300
Sys Low
Mean Low
ART
Ao
FAP
BAP
UAP
Dia Low
0 to (high limit – 2)
1
Sys High
Mean High
Dia High
(low limit + 2) to 120
Sys Low
Mean Low
PA
Dia Low
-6 to (high limit – 2)
1
Mean High (low limit + 2) to 40
CVP, LAP
RA P, IC P,
UVP
Mean Low -10 to (high limit – 2)
1
Sys High
Mean High
Dia High
(low limit + 2) to 300
Sys Low
Mean Low
P1 to P4
Dia Low
-50 to (high limit – 2)
1
A.5.8 C.O.
Measurement method Thermodilution method
C.O.: 0.1 to 20 L/min
TB:
23 to 43°C
Measurement range
TI:
0 to 27°C
Resolution
C.O.:
TB, TI:
0.1 L/min
0.1°C
Accuracy
C.O.:
TB, TI:
±5% or ±0.1 L /min, whichever is greater
±0.1°C (without sensor)
Alarm range BT:
23 to 43°C