C-6
C.2.4 SpO2
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Alarm * * On
Alm Lev * * Med
SpO2 High * * Adu, Ped:
Neo:
100
95
SpO2 Low * * Adu, Ped, Neo: 90
Desat Limit * * Adu, Ped, Neo: 80
Sweep * * 25 mm/s
NIBP Simul * Off
Pitch Tone * Mode 1
Sensivity * * Med
Sensivity (Masimo) * * Normal
Averaging (Masimo, Nellcor) * * 8 s
Sat-Seconds (Nellcor) * * 0 s
C.2.5 Temp
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Alarm * * On
Alm Lev * * Med
Temp Unit * * ºC
Adu 39.0
Ped 39.0
T1/T2 High (ºC)
Neo
* *
39.0
Adu 36.0 35.0 36.0
Ped 36.0 35.0 36.0
T1/T2 Low (ºC)
Neo
* *
36.0 35.0 36.0
Adu 2.0
Ped 2.0
TD High (ºC)
Neo
* *
2.0