C-7
C.2.6 NIBP
O.M Item Name
C M
General OR ICU NICU CCU User
Defaults
Alarm * * On
Alm Lev * * Med
Alm Source * * All
Measure Mode * * Manual Auto Manual Auto
Interval * * 15 min 5 min 15 min 30 min 15 min
Press. Unit * * mmHg
Display NIBP * * Single-group Multi-group Single-group
PR Display * Off
Adu 80
Ped 60
Cuff Press
(mmHg)
Neo
* *
40
Adu 160
Ped 140
Inflation
Pressure
(mmHg)
Neo
* *
90
Alarm Limits Setup
Adu 160 180 160
Ped 120 140 120
Sys High
(mmHg)
Neo
* *
90
Adu 90 70 90
Ped 70 60 70
Sys Low
(mmHg)
Neo
* *
40
Adu 110 120 110
Ped 90
Mean High
(mmHg)
Neo
* *
70
Adu 60
Ped 50
Mean Low
(mmHg)
Neo
* *
25
Adu 90
Ped 70
Dia High
(mmHg)
Neo
* *
60
Adu 50
Ped 40
Dia Low
(mmHg)
Neo
* *
20