EasyManua.ls Logo

Mindray iPM-9800 - Page 274

Mindray iPM-9800
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
C-8
C.2.7 IBP
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Alarm * * On
Alm Lev * * Med
P1 Measure * * All
P2 Measure * * Mean
P3 Measure * * All
P4 Measure * * Mean
Alarm Source * * All
Press. Unit * * mmHg
Response * * Med
Sweep * * 25 mm/s
Auto Scaling * * Off
Filter * 12.5 Hz
Wave Overlapping * * Off
Art, Ao, UAP, BAP, FAP Setup
Upper Scale * * 150
Middle Scale * * 75
Lower Scale * * 0
Adu 160 180 160
Ped 120 140 120
Sys High
(mmHg)
Neo
* *
90
Adu 90 70 90
Ped 70 60 70
Sys Low
(mmHg)
Neo
* *
55
Adu 110 120 110
Ped 90
Mean High
(mmHg)
Neo
* *
70
Adu 70 60 70
Ped 50
Mean Low
(mmHg)
Neo
* *
35
Adu 90
Ped 70
Dia High
(mmHg)
Neo
* *
60
Adu 50
Ped 40
Dia Low
(mmHg)
Neo
* *
20