Loading...
C-11
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Microstream CO2 Setup
Humidity Compen * Off
Max Hold * * 20 s
Auto Standby (min) * * 0
Mainstream CO2 Setup
Max Hold * * 10 s
O2 Compen * Off
Balance Gas * Room Air
AG Compen * 0
Alarm Limits Setup
EtCO2 High (mmHg) * * Adu, Ped:
Neo:
50
45
EtCO2 Low (mmHg) * * Adu:
Ped:
Neo:
15
20
30
FiCO2 High (mmHg) * * Adu, Ped, Neo: 4
awRR High * * Adu, Ped:
Neo:
30
100
awRR Low * * Adu, Ped:
Neo:
8
30
Apnea Time * * 20 s 15 s 20 s