C-12
C.2.10 AG
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Alarm * * On
Alm Lev * * Med
Sweep * * 6.25 mm/s
O2 Compen * Off
Operating Mode * * Measure
Flow Rate * * Adu, Ped:
Neo:
120 ml/min
70 ml/min
Auto Standby * * Off
Apnea Time * * 20 s
CO2 Setup
CO2 Unit * * mmHg
Wave Type * * Draw
Scale * * when unit is mmHg:
when unit is % or KPa:
50
7.0
EtCO2 High * * Adu, Ped:
Neo:
50
45
EtCO2 Low * * Adu:
Ped:
Neo:
15
20
30
FiCO2 High * * 4
FiCO2 Low * * 0
awRR High * * Adu, Ped:
Neo:
30
100
awRR Low * * Adu, Ped:
Neo:
8
30