D-6
Measurement Alarm message L I Cause and solution
NIBP Illegally Reset L C An illegal reset occurred during NIBP
measurement. Check if the airway is occluded.
IBP YYSensor Off
(YY represents an
IBP label.)
LA
C.O. TB Sensor Off L A
Check the sensor connection and reconnect the
sensor.
CO2 Sensor High
Temp
L C Check, stop using or replace the sensor.
CO2 Sensor Low
Temp
L C Check, stop using or replace the sensor.
CO2 Airway High
Press.
LC
CO2 Airway Low
Press.
LC
An error occurred in the airway pressure. Check
the patient connection and patient circuit, and then
restart the monitor.
CO2 High
Barometric
LC
CO2 Low
Barometric
LC
Check the CO2 connections, make sure that the
monitor application site meets the requirements,
and check for special sources that affect the
ambient pressure. Restart the monitor.
CO2 FilterLine
Occluded
L C The airway or watertrap was occluded. Check the
airway and remove the occlusion.
CO2 No Watertrap L B Check the watertrap connections.
CO2 Check Adapter L C There is a problem with the airway adapter.
Check, clean or replace the adapter.
CO2 FilterLine Err L C Check if there is a leak in the CO
2
sample line or
the CO
2
sample line has been occluded.
CO2 Zero Failed L C Check the CO
2
connections. After the sensor’s
temperature becomes stabilized, perform a zero
calibration again.
CO2 Check Cal. L C Perform a calibration.
CO2 Check Airway L C An error occurred to the airway.
CO2 No Filterline L A Make sure that the filterline is connected.
CO2 Main Board
Err
HC
CO2 Replace
Scrubber&Pump
LC
CO2 15V Overrange H C
CO2 Hardware Err H C
There is a problem with the CO
2
module. Restart
the monitor.
CO
2
CO2 User Cal Fail L C Re-calibrate the AG module.