D-7
Measurement Alarm message L I Cause and solution
CO2 Check Sensor L C The CO2 sensor appears abnormal. Check or
replace the sensor.
CO2 No Sensor L A Make sure that the CO2 sensor is connected.
CO2 Temp
Overrange
L C The module detects that the temperature has
exceeded the specified range. Please move away
the source of heat or wait for the module to
recover.
AG No Watertrap L B Check the connections of the watertrap and
re-connect it.
AG Change
Watertrap
L C Wait until the change is completed.
AG Watertrap Type
Wrong
L C Make sure that a correct watertrap has been used.
YY Accuracy
Unspecified
(YY represents O2,
N2O, CO2, Enf, Iso,
Sev, Hal, Des, or
awRR (AG)
L C The measured value has exceeded the specified
accuracy range.
AG Hardware
Mulfunction
H C Stop using the AG module and contact your
service personnel.
AG Airway
Occluded
L C Check the airway and remove the occlusion.
AG Zero Failed L C Restart the monitor, and then perform a zero
calibration again.
AG Cal. Failed L C Re-calibrate the AG module.
Anesthetic Mixture L C The system has detected more than one anesthesia
gas. Check the ventilated anesthesia gas.
O2 Sensor Err L C The O2 sensor appears abnormal. Contact your
service personnel.
AG
AG Limit Error L C AG violated set alarm limits seriously and
measurements cannot be performed. Contact your
service personnel.
12V Too High H C
12V Too Low H C
3.3V Too High H C
3.3V Too Low H C
There is a problem with the system power supply.
Restart the monitor.
Power
Low Battery M C Connect the monitor to an AC power source and