D-8
Measurement Alarm message L I Cause and solution
Battery Depleted HC
allow the batteries to charge.
Battery Mismatched M C The two batteries have different charge capacity,
or the batteries unspecified have been used, or
there is a problem with the batteries. Make sure
that correct batteries are used and the batteries are
not damaged, or replace the batteries.
RT Clock Need
Reset
L C Re-set the system time and restart the monitor.
RT Clock Not Exist H C Contact your service personnel.
Power Board
Comm. Stop
H C Contact your service personnel.
Recorder SelfTest
Err
LC
Recorder Comm Err L C
Stop the recording and restart the monitor.
Recorder
Recorder Head Hot L C The recorder has been working for too long time.
Stop the recording and resume the recording till
the recorder’s printhead cools down.
IP Address Conflict L C Check the monitor’s IP address and re-set it.
Loading User
Config. Failed.
LC
Loading Factory
Config. Failed.
LC
The system re-loads configuration automatically.
If the re-leading always failed, contact your
service personnel.
CF Storage Card
Err!
M C An error occurred when the CF storage card reads
and writes. Contact your service personnel.
System
CF Card Space Low L B CF card space is lower than 40 M.