12 - 18 Operator’s Manual
12 Probes and Biopsy
L12-3VNs Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
L12-3RCs Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
L14-6Ns Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
L20-5s Operating 0°C ~ 35°C 15% ~ 80% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 60°C 15% ~ 90% 500 hPa ~ 1060 hPa
7LT4s Operating 0°C ~ 40°C 25% ~ 90% 550 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 60°C 10% ~ 95% 500 hPa ~ 1060 hPa
P4-2s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
P10-4s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
SP5-1Ns Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
L12-4s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
V11-3Ws Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
SP5-1s Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
P7-3s Operating 10°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
SC6-1s Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
Table 12-3 Probe Environmental Conditions
Probe
Model
Conditions Ambient
temperature
Relative humidity
(no condensation)
Atmospheric
pressure