12 Probes and Biopsy
Operator’s Manual 12 - 17
Compatible Cleaners and Disinfectants
For the detailed information about cleaners and disinfectants, see Mindray Transducer Disinfectant
Recommendation.
12.1.7 Cleaning the probe cable and connector
Do not use cloth with water to clean the probe connector.
Perform the following procedure:
1. Wipe out the dust attached to surface of probe connector and cable.
2. Use soft brush to brush the dust inside probe connector gently.
3. Remained stain or dust attached to surface of cable or surface of connector should be washed
out by cloth with a little soapy water, and then air-dry.
12.1.8 Probe Environmental Conditions
Table 12-3 Probe Environmental Conditions
Probe
Model
Conditions Ambient
temperature
Relative humidity
(no condensation)
Atmospheric
pressure
C4-1s Operating 0°C ~ 35°C 15% ~ 90% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 60°C 15% ~ 90% 700 hPa ~ 1060 hPa
C5-1s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
C5-2s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
C11-3s Operating 0°C ~ 40°C 30% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 30% ~ 95% 700 hPa ~ 1060 hPa
SC5-1Ns Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
L9-3s Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa
L11-3VNs Operating 0°C ~ 40°C 20% ~ 85% 700 hPa ~ 1060 hPa
Storage and
transportation
-20°C ~ 55°C 20% ~ 95% 700 hPa ~ 1060 hPa