2 - 2 Operator’s Manual
2 System Overview
According to the degree of safety of application in the presence of a FLAMMABLE
ANESTHETIC MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE
EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANESTHETIC
MIXTURE WITH AIR OR WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation:
Continuous operation
According to the installation and use:
Portable equipment
Mobile equipment (when the system is installed on the trolley)
Does the equipment has any defibrillation-proof applied parts:
Non-defibrillation-proof applied part
Permanently installed equipment or non-permanently installed equipment:
Non-permanently installed equipment
2.4 Product Specifications
NOTE:
The functions described in the operators manual may vary depending on the specific system
purchased.
2.4.1 Power supply
Voltage: 100-240V~; 19Vdc (DCU direct input)
Frequency: 50/60Hz
Power input: 2.0A (Power adapter); 5.4-2.6A (Trolley)
Fuse: T5AL, 250VAC (Power adapter)
•Battery: 14.8Vdc
2.4.2 Environmental Conditions
Operating conditions
Ambient temperature: 0 °C ~ 40 °C
Relative humidity: 30% ~ 85% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
Storage and transportation conditions
Ambient temperature: -20 °C ~ 55 °C
Relative humidity: 20% ~ 95% (no condensation)
Atmospheric pressure: 700 hPa ~ 1060 hPa
WARNING
Do not use this system in conditions other than those specified.