12 Probes and Biopsy
Operator’s Manual 12 - 53
6. Rinse the needle-guided bracket thoroughly by using a large amount of distilled or softened
water (about 2 gallons) at room temperature for about 30 s to remove the residual dirt and
cleaning solvent. Repeat the operation twice.
7. Wipe away the water on the needle-guided bracket by using a piece of disposable lint-free soft
cloth.
8. Inspect the needle-guided bracket. If visible dirt still exists, repeat the preceding steps to wash
the bracket until it is all clean.
Sterilization with a sterilant
Perform the following procedure:
1. Wear a pair of gloves to prevent infection.
2. Clean thoroughly in accordance with the cleaning procedure before sterilization.
3. Prepare a sterilant by using sterile distilled water when necessary.
4. Immerse the needle-guided bracket fully in the sterilant and shake the bracket appropriately to
remove any bubbles on the surface. Use a syringe to draw an appropriate amount of sterilant
and inject the sterilant into the hole to remove the bubbles inside the hole if necessary.
For details about the immersion duration, see the operator’s manual provided by the
manufacturer.
5. After sterilization, wash the needle-guided bracket thoroughly by using a large amount of
sterile distilled water (about 2 gallons) at room temperature for about 30s to remove the
residual sterilant. Repeat the operation twice.
6. Dry the needle-guided bracket by using a piece of sterile disposable lint-free soft cloth.
7. Store the needle-guided bracket in a cool, clean and dry environment.
High-pressure steam sterilization
High-pressure steam sterilization is preferred for the metal guided brackets. And the sterilizer and
accessories should be cleared by FDA for the intended sterilization cycle.
Perform the following procedure:
1. Wear a pair of gloves to prevent infection.
2. Clean thoroughly in accordance with the cleaning procedure before sterilization.
3. Package the needle-guided bracket in accordance with the sterilization requirements of
surgical instruments using the sterilization wrap or pouch cleared by FDA.
Table 12-5 Recommended sterilization solution
Chemical name Trade name Procedures
Glutaraldehyde (2.4%) Cidex Activated Dialdehyde
Solution
(applicable for FDA region
only)
Refer to the instructions
provided by the solution
manufacturer for details.
22% Hydrogen Peroxide
4.5% Peroxyacetic Acid
Minncare liquid disinfectant
(applicable for Canada only)
Refer to the instructions
provided by the solution
manufacturer for details.
Glutaraldehyde (2.6%) Metricide Refer to the instructions
provided by the solution
manufacturer for details.