14 System Maintenance
Operator’s Manual 14 - 3
NOTE:
Use a soft brush to brush away dust attached to all visible sockets or interfaces (such as probe
sockets, sockets or interfaces in the IO panel and power supply panel). Do not use a cloth and water.
Cleaning the dust-proof cover
CAUTION
Please clean all the system’s dustproof covers periodically (once per month).
Failure to do so may result in system damage. Cleaning times can be
increased when the system is used in the open air or somewhere more dusty.
Perform the following procedure:
1. Press the clip of the dust-proof cover and unfold it outwards (shown in the figure below).
2. To remove the fixing frame of the dust-proof cover and the cover itself, press two clips on the
fixing frame of the dust-proof cover towards arrow’s direction (shown in the figure below).
3. Clean the dust-proof mesh with water or soft brush.
Clip
Clip
Fixing frame