4 - 12 Operator’s Manual
4 Setup
2. Select the imported file.
3. Tap [OK], a progress bar will appear and the setup data is imported to the specified path.
4. To restore the factory setup data, tap [Load Factory] on the right side of the screen.
Probe Check
This function is used to check the condition of transducer elements to evaluate the probe
performance.
Before performing the probe check, ensure the probe head must be clean and the probe is in non-
scanning state.
Tap [Probe Check] to enter the Probe Check screen:
• If a transducer element is in malfunction, it is displayed as a red spot.
• If a transducer element functions well, it is displayed as a green spot.
You can export the result image to the external device.
Other Settings
Other preset settings are described as follows:
Item Description
Load Factory Local factory default settings.
Setup Show Metrics Logs Show the metrics logs, such as login count, total usage time,
user metrics details, exam mode metrics details or probes.
Extract Preset Data /
Export Log Export the log.
Self Test Perform system self-test and restart the machine.
Recover To recover the system.
Enter Windows For password, please contact the service engineer.
Needles List Replace Update the list of needles that support eSpacial Navi function.
NOTE:
The list of needles must be release version provided by
Mindray.
Needles List Recover Restore the list of needles that support eSpacial Navi function
to the last version.
Needles List Factory Restore the list of needles that support eSpacial Navi function
to the factory version.
Fast Startup To speed up the startup process.
NOTE:
After installing McAfee, please disable the Fast Startup
function to avoid McAfee failure.
Transducer Tracking Enable the transducer and patient tracking function:
• Find transducers used with a patient by patient ID.
• Find patients who used a transducer by transducer SN.
After enabling the tracking function, you can set the maximum
number of days the data can be retained.