7-16 Structure and Assembly/Disassembly
Figure 7-13 Disassemble the DC-DC Power Board (1)
2. Remove the plug of fan cable.
Figure 7-14 Disassemble the DC-DC Power Board (2)
3. After removing the connecting plug between DC-DC power board and AC-DC power board,
you could take out the DC-DC power board.
Figure 7-15 Disassemble the DC-DC Power Board (3)
7.4.2.3 AC-DC Power Board
1. Remove panhead screws with washers M3X6 (5) used to fix the AC_DC power board.