Software Upgrade and Maintenance 6-5
Other steps are the same as method “2”.
When enter 3 and press <Enter>, if the model type is the same as service restoration package (for
example, human model with human restoration package or vet model with vet restoration package),
the conversion will be failed .System installation will stop without any change.
6. After the system has been installed successfully, the following information will display on the
screen, press <Enter> to restart the system.
7. After the OS is restored, Doppler restoration starts automatically. The process of restoring refer
to 6.2.2 Doppler restoration (from the 5
th
to the 8
th
step)
6.2.1.3 File Processing during Restoration
During the OS installation procedure, the machine will back up the key file in the HDD
automatically, see below:
Select the choice 1:
1. Back up the option key file of the C drive to D:/mnt/bkcfg (temporary catalogue)
2. Format the C drive and restore the OS.
3. Copy the option Key file to the previous catalogue in C drive, and the temp file will be
deleted.
Select the choice 2:
1. Back up the option key file of the C drive to U:/bkcfg (temporary catalogue)
2. Format the whole HDD and restore the OS.
3. Copy the option Key file to the previous catalogue in C drive, and the temporary file in U
disk will be deleted.
In the OS with previous configuration data, if the ultrasound system is restored by package
without configuration data, the product configuration and the option state will take effect as the
configuration data in OS.