Software Upgrade and Maintenance 6-7
5. The system then will ask to restart the system. Click [OK] to enter restore procedure, click
[Cancel] will cancel the operation and returns to Doppler.
6. During the course, when the restoration is almost done, the following prompts will display:
Enter “Y” or “y” to delete the data in D:\z5\Preset\Enter "N" and the system will not delete data.
NOTE:
The old user preset data may be not compatible with the restored system, if
this happen, the operator must select “Y” when the system asks whether to delete
user preset data or not, otherwise, the system can’t be restarted normally.
7. After the whole system is restored, if there is only green shutdown prompts displayed, the
restore process is normal, and press <Enter> to quit.