Command name Command Number Return command set value
Decimal point display position 71 01
0 : No decimal point, 1 : 1234.5, 2 : 123.45
3 : 12.345, 4 : 1.2345
A/D sampling rate 71 02
0 : 50 times/s, 1 : 100 times/s, 2 : 250 times/s
3 : 500 times/s
Overload display condition 71 03
0 : When exceeding | Weighing capacity + 9D |
1 : When exceeding ± 110 of weighing capacity
2 : When exceeding -20D ~ |Weighing capacity
+9D |
Comparative operation display
interlocked
71 04 0 : Invalid, 1 : Valid
Unit 71 05
0 : No unit, 1 : g, 2 : kg, 3 : t, 4 : lb, 5 : N, 6 : kN,
7 : Pa, 8 : kPa, 9 : MPa
ZERO set operation condition 71 10
0 : Operation in stable condition,
1 : Unconditionally operation
ZERO set effective range 71 11 00 ~ 30 (Unit : ± 1 % of weighing capacity)
Target of ZERO tracking 71 12
0 : Gross and Net value (against load displ
ay)
1 : Gross value (only in the display of gross weight)
ZERO tracking data width 71 13 000 ~ 999 (Unit : 0.1D)
ZERO tracking time width 71 14 00 ~ 99 (Unit : 0.1 s)
Power-on-ZERO operation 71 15 0 : Invalid, 1 : Valid
Operative condition of tare weight
cancellation
71 16
0 : Operation during stable with 0 < Gross value
<= Weighing capacity.
1 : Operation always with 0 < Gross value
<= Weighing capacity.
2 : O
peration during stable with Gross value
<= | Weighing capacity |
3 : Operation always with Gross value <= | Weighing
capacity |.
Change of data stored place 71 17
0 : Internal RAM, 1 : EEPROM
10
0
digit : ZERO set data
10
1
digit : Tare weight cancellation data
Detecting stability while setting 71 20 0 : Invalid, 1 : Valid
Setting method of gravity
acceleration compensation value
71 25
0 : Set the district nu
mber
1 : Set the numeric value of gravity acceleration
District number of using place 71 26 01 ~ 16 (Unit : District)
District number of calibration place 71 27 01 ~ 16 (Unit : District)
Gravity acceleration value of using
place
71 28 9.000 ~ 9.999 (Unit : m/s
2
)
Gravity acceleration value of
calibration place
71 29 9.000 ~ 9.999 (Unit : m/s
2
)
Calibration data (for reference)
- Scale interval
71 90
0 : 1 scale interva
l, 1 : 2 scale interval,
2 : 5 scale interval, 3 : 10 scale interval,
4 : 20 scale interval, 5 : 50 scale interval,
Calibration data (for reference)
- Weighing capacity
71 91 00000 ~ 99999
Calibration data (for reference)
- Mass of weight
71 92 00000 ~ 99999
Calibration data (for reference)
- ZERO mV/V value
71 93 - 3.1000 ~ 3.1000
Calibration data (for reference)
- SPAN mV/V value
71 94 - 3.1000 ~ 3.1000
Stabi
lity detecting time width while
calibration
71 97 00 ~ 99 (Unit : 0.1 s)
• Please change [7*] in above mentioned command part to [8*] at the time of writing.
• The number of [90] ~ [94] is only reading because of the items relating to the
calibration.
123