5-8 Calibration and Adjustment
9. Verify that the calibration weight is set to 50 and click on Calibrate.
! FmXP will prompt you to place the weight on the scale.
10. Using gloves or tweezers, place the 50 g calibration weight on the pedestal.
11. Replace the scale cover and click
OK.
! FmXP will display the following message (
Figure 5-9):
Figure 5-9 FmXP Calibrating Scale Message
12. Upon completion, FmXP displays the following message:
Figure 5-10 FmXP Calibration Complete Message
13. Remove the weight from the scale and click OK.