EasyManua.ls Logo

Olympus OmniScan MX2 - Acquisition Specifications; Data Specifications; J.2 Acquisition Specifications; Table 57 Acquisition Specifications

Olympus OmniScan MX2
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-20015-01EN [U8778402], Rev. J, January 2014
OMNI-M-PA3232 Module Specifications
165
J.2 Acquisition Specifications
This section details the acquisition specifications of frequency, data display, and
synchronization (see Table 57 on page 165).
J.3 Data Specifications
This section details the data specifications of processing, TCG, and storage (see
Table 58 on page 165).
Table 57 Acquisition specifications
Frequency
Digitizing frequency 100 MHz (10 bits)
A-scan (acquisition
mode)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Maximum pulsing rate Up to 10 kHz (C-scan)
Depth in material 59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
Refresh rate A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Envelope (echo-dynamic
mode)
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
On internal clock 1 Hz to 10 kHz
External pace Yes
On encoder On 2 axes: from 1 step to 65536 steps
Table 58 Data specifications
Processing
Number of data points Up to 8000
Real-time averaging 2, 4, 8, and 16
Rectification RF, full wave, half wave +, half wave

Table of Contents

Other manuals for Olympus OmniScan MX2

Related product manuals