OMICRON 173
Technical data
12.8 Environmental conditions
12.9 Mechanical data
12.10 Standards
Table 12-11: Environmental conditions
Characteristic Rating
Operating temperature –10 … +50°C (14 … 122°F)
Storage and transportation –25 … +70°C (–13 … 158°F)
Max. altitude for operation 2000m
Table 12-12: Mechanical data
Characteristic Rating
Weight < 8kg (17.6lbs) without accessories
Dimensions W x H x D 360 x 285 x 145mm (14.2 x 11.2 x 5.7")
Table 12-13: Standards
EMC, safety
EMC
IEC/EN 61326-1 (industrial electromagnetic environment)
FCC subpart B of part 15, class A
Safety IEC/EN/UL 61010-1
Other
Shock IEC/EN 60068-2-27 (15 g/11 ms, half-sinusoid, 3 shocks in each axis)
Vibration
IEC/EN 60068-2-6 (frequency range 10 Hz…150 Hz, acceleration 2 g
continuous (20 m/s
2
/65 ft/s
2
), 20 cycles per axis)
Humidity
IEC/EN 60068-2-78 (5%…95% relative humidity, no condensation),
tested at 40 ºC/104 ºF for 48 hours
Protection class IP20 according to EN 60529