CT Analyzer User Manual
54 OMICRON
Figure 5-2: Short guide for testing in Advanced CT Test mode (2)
Test execution:
Burden test
Test finished message
-> 5.2.3, step 11. on page 59
Disconnect or perform new test
-> 5.2.5 on page 61
Burden test card
enabled?
Connect CT Analyzer
for burden test
-> 3.4.2 on page 29
Start Test
soft key
Skip Test
soft key
Cancel Test
soft key
Connect CT Analyzer
for CT test
-> 3.4.1 on page 27
Prim.
Winding Resistance
measurement
enabled?
Connect CT Analyzer for
primary winding resistance
measurement
-> 3.4.3 on page 30
Start Test
soft key
Skip Test
soft key
Cancel Test
soft key
View results
-> 5.2.4 on page 60
Start Test
soft key
Skip Test
soft key
Cancel Test
soft key
Test execution:
CT test
from previous page
Yes
No
Yes
No
Test execution:
Prim. winding resistance
measurement
Residual Magnetism
measurement if enabled
-> 8.4 on page 109

Table of Contents

Other manuals for Omicron CT SB2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CT SB2 and is the answer not in the manual?

Omicron CT SB2 Specifications

General IconGeneral
BrandOmicron
ModelCT SB2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals