EasyManua.ls Logo

ProMinent Splash Control Pro 2 - Measuring Input Connections

ProMinent Splash Control Pro 2
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A2997
Fig. 8: Primary power supply
3. Guide the cable through a suitable cable gland.
4. Connect the phase to L1 and the neutral conductor to the N
of the main terminal X1.
5. Use an M4 eyelet clamp to connect the earth to the PL1 bolt.
6. Tighten the cable gland to make a seal.
4.3.7 Measuring input connections
The device has 10 inputs:
n 2 potentiometric inputs PI1 & PI2 for pH and ORP sensors,
n 2 inputs AI1 & AI2, 4 ... 20 mA, insulated for temperature,
chlorine or bromine measurement,
n 2 inputs AI3 & AI4, 4 ... 20 mA, uninsulated for temperature,
chlorine or bromine measurement,
n 4 digital inputs DI1 ... DI4 for sensors.
A2998
Fig. 9: a) Potentiometric inputs PI1 & PI2
The controller has two potentiometric inputs to which a pH or ORP
sensor can be connected.
Tab. 2: The supported sensors are defined as follows (pH):
pH value 1...12 pH value 0...14 Customer*
pH value (input PI1 or PI2) -• -• -•
(*): The customer’s scale can be defined between -1 ... 15 pH.
Tab. 3: The supported sensors are defined as follows (ORP):
0...1000 mV +/- 1000 mV Customer*
ORP value (input PI1 or PI2) -• -• -•
(*): The customer’s scale can be defined between -1000 ... 1000 mV.
Installation and assembly
20

Table of Contents

Related product manuals