Page 75
To program the Ultra Twin for Example 1 V-Notch Weir by using the
Quick Setup menu proceed as follows.
If required access the Program Mode
Key in the passcode 1997 and press ENTER
At the Quick Setup menu press ENTER and as prompted, by the questions,
select the relevant option and ENTER.
Question
Option
Application
3 = Flow
PMD Type
1 = Exponent
Exponent
6 = V notch.
Calculation
2 = Ratiometric.
No. of Alarms
0 = No Alarms
Xducer (P1-101)
1 = dB Mach3
Volume Units (P1-706)
6 = Mil USG
Time Units (P1-707)
4 = Per Day
Measnt. Units (P*104)
5 = inches
Empty Level (P1-105)
40 inches
Minimum Head (P1-703)
0.000 inches
Maximum Head (P1-704)
15 inches
Total Alloc. (P1-824)
1 = Point 1
Aux. Mode (P1-815)
2 = Level
Aux. Source (P1-816)
0 = Off
Total Multiplier (P1-823)
1 = /1000
Maximum Flow (P1-705)
2.2
Programming is now complete and the unit can now be returned to the run
mode, press CANCEL until Run Mode? Is displayed on the LCD press
ENTER, and the Ultra Twin will return to Run Mode.