RIGOL 2 Command System
2-306 DS2000E Programming Guide
:TRIGger:USB:MLEVel
:TRIGger:USB:PLEVel
Syntax
:TRIGger:USB:MLEVel <level>
:TRIGger:USB:MLEVel?
:TRIGger:USB:PLEVel <level>
:TRIGger:USB:PLEVel?
Description
Set or query the trigger level of USB trigger when the channel source of the D- data line is CHANnel1 or
CHANnel2. The unit is the same with the current amplitude unit.
Set or query the trigger level of USB trigger when the channel source of the D+ data line is CHANnel1 or
CHANnel2. The unit is the same with the current amplitude unit.
Parameter
Name Type Range Default
<level> Real (-5 x VerticalScale - OFFSet) to (+5 x VerticalScale - OFFSet) 0
Note: For the VerticalScale, refer to the :CHANnel<n>:SCALe command. For the OFFSet, refer to
the
:CHANnel<n>:OFFSet command.
Explanation
To set the channel sources of the D- and D+ data line of USB trigger, please refer to
the
:TRIGger:USB:DMINus and :TRIGger:USB:DPLus commands respectively.
Return Format
The query returns the trigger level of USB trigger when the channel source of the D- or D+ data line is
CHANnel1 or CHANnel2 in scientific notation.