2 Command System RIGOL
DS2000E Programming Guide
2-41
:BUS<n>:CAN:BAUD
Syntax
:BUS<n>:CAN:BAUD <baud>
:BUS<n>:CAN:BAUD?
Description
Set or query the signal rate of CAN decoding on the specified bus. The default unit is b/s.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<baud> Discrete {100000|125000|250000|400000|500000|800000|1000000|USER} 500000
Note: When the signal rate is set to USER, you can use the :BUS<n>:CAN:BUSer command to set a user-defined
signal rate.
Return Format
The query returns the signal rate of CAN decoding on the specified bus in integer or returns USER.
:BUS<n>:CAN:BUSer
Syntax
:BUS<n>:CAN:BUSer <baud>
:BUS<n>:CAN:BUSer?
Description
Set or query the user-defined signal rate of CAN decoding on the specified bus. The default unit is b/s.
Parameter
Name Type Range Default
<n> Discrete {1|2} --
<baud> Integer 10000 to 1000000 500000
Return Format
The query returns the user-defined signal rate of CAN decoding on the specified bus in integer.
:BUS<n>:CAN:SPOint
Syntax
:BUS<n>:CAN:SPOint <spo>
:BUS<n>:CAN:SPOint?