Operators Manual · Version 1.0 C-39
URISYS 2400 6 Special operations
Calibration
used for measurement (B = blue 470 nm, G = green 555 nm, O = orange
620 nm).
In case of Calibr. Failed, the values out of range are displayed with an
asterisk on the screen and the printout.
Insert the new calibration strip correctly and press <Calibrate>, followed by
<Execute> in the confirmation screen.
a To print calibration values
1 Press <Print> to print the calibration values. A confirmation window
appears.
2 Press <Execute> to confirm. The calibration data are printed.
3 Press <Home> to return to the [Routine Monitor] screen.
Figure C-13 Photometer Calibration screen
If the calibration failed, an alarm message appears. Repeat the calibration with
another calibration strip from the vial.