EasyManua.ls Logo

Rohde & Schwarz R&S RTB2004 - Page 269

Rohde & Schwarz R&S RTB2004
509 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
RTB2000
261User Manual 1333.1611.02 ─ 04
Example:
ACQ:POIN:AUT ON
TIM:SCAL 1e-9
ACQ:POIN?;:SYST:ERR:ALL?
-> received 10000;0,"No error"
TIM:SCAL 5e-3
ACQ:POIN?;:SYST:ERR:ALL?
-> received 20000000;0,"No error"
ACQuire:POINts[:VALue]
Defines a record length value, the number of recorded waveform points in a segment.
The command turns ACQuire:POINts:AUTomatic OFF.
If ACQuire:POINts:AUTomatic is turned ON, the query ACQuire:POINts? returns
the automatically set record length.
Each predefined record length corresponds to a maximum number of history seg-
ments, which are stored in the instrument's memory. If option R&S RTB-K15 is instal-
led, you can display the history segments.
Available record length values are:
10 kSa (13107 history segments)
20 kSa (6553 history segments)
50 kSa (2621 history segments)
100 kSa (1456 history segments)
200 kSa (771 history segments)
500 kSa (319 history segments)
1 MSa (159 history segments)
2 MSa (80 history segments)
5 MSa (32 history segments)
10 MSa (16 history segments)
20 MSa (8 history segments)
Parameters:
<RecordLength> Record length in Sa.
If the entered value differs from the predefined values, the
instrument sets the closest value.
CHANnel<m>:TYPE <DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform
points with lower sample rate.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Waveform Setup