Remote Control Commands
R&S
®
RTB2000
262User Manual 1333.1611.02 ─ 04
Parameters:
<DecimationMode> SAMPle | PDETect | HRESolution
SAMPle
Input data is acquired with a sample rate which is aligned to the
time base (horizontal scale) and the record length.
PDETect
Peak Detect: the minimum and the maximum of n samples in a
sample interval are recorded as waveform points.
HRESolution
High resolution: The average of n sample points is recorded as
waveform point.
*RST: SAMPle
ACQuire:TYPE <AcquisitionType>
Sets the type of the aquisition mode.
Parameters:
<AcquisitionType> REFresh | AVERage | ENVelope
REFresh
The aquisitions are displayed as they are done.
AVERage
The aquisitions are averaged.
ENVelope
The envelope of a repetitive signal is shown, representing the
borders in which the signal occurs.
CHANnel<m>:ARIThmetics <TrArith>
Selects the method to build the resulting waveform from several consecutive acquisi-
tions of the signal.
Suffix:
<m>
.
1..4
The command affects all channels regardless of the indicated
channel number. The suffix can be omitted.
Waveform Setup