Remote Control Commands
R&S
®
RTB2000
263User Manual 1333.1611.02 ─ 04
Parameters:
<TrArith> OFF | ENVelope | AVERage
OFF
The data of the current acquisition is recorded according to the
decimation settings.
ENVelope
Detects the minimum and maximum values in an sample interval
over a number of acquisitions.
AVERage
Calculates the average from the data of the current acquisition
and a number of acquisitions before. The number of used
acquisitions is set with ACQuire:AVERage:COUNt.
*RST:
OFF
ACQuire:PEAKdetect <PeakDetect>
Enables or disables the peak detect acquisition mode.
You can use this command alternatively to CHANnel<m>:TYPE.
Parameters:
<PeakDetect> AUTO | OFF
*RST: OFF
ACQuire:HRESolution <HighRes>
Enables or disables the high resolution acquisition mode.
You can use this command alternatively to CHANnel<m>:TYPE.
Parameters:
<HighRes> AUTO | OFF
*RST: OFF
ACQuire:NSINgle:COUNt <NSingleCount>
Sets the number of waveforms acquired with RUNSingle.
Parameters:
<NSingleCount> Number of waveforms.
Range: 1 to maximum number that depends on the record
length.
*RST: 1
ACQuire:AVERage:COUNt
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.