Remote Control Commands
R&S
®
RTB2000
264User Manual 1333.1611.02 ─ 04
Parameters:
<AverageCount> Range: 2 to 100,000
*RST: 2
ACQuire:AVERage:RESet
Deletes the waveform and restarts the average calculation.
Usage: Event
ACQuire:AVERage:COMPlete?
Returns the state of averaging.
Return values:
<AverageComplete> 0 | 1
0
The number of acquired waveforms is less than the number
required for average calculation. See ACQuire:AVERage:
COUNt on page 263.
1
The instrument acquired a sufficient number of waveforms to
determine the average.
Usage: Query only
TIMebase:ROLL:ENABle <Roll>
Enables the roll mode.
Parameters:
<Roll> ON | OFF
*RST: OFF
ACQuire:INTerpolate <InterpolationType>
Defines the interpolation mode.
See also: "Interpolation" on page 52
Parameters:
<InterpolationType> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & hold causes a histogram-like interpolation.
*RST: SINX