Remote Control Commands
R&S
®
RTB2000
388User Manual 1333.1611.02 ─ 04
Suffix:
<b>
.
1 | 2
<n> *
Selects the frame.
Return values:
<DataWords> Comma-separated list of decimal values of the data bytes.
Example:
BUS:I2C:FRAM2:DATA?
returns four data bytes:
-> 69,158,174,161
Usage: Query only
BUS<b>:I2C:FRAMe<n>:STATus?
Returns the overall state of the frame.
Suffix:
<b>
.
1 | 2
<n> *
Selects the frame.
Return values:
<State> INComplete | OK | UNEXpstop | INSufficient | ADDifferent
INComplete
The frame is not completely contained in the acquisition.
Usage: Query only
BUS<b>:I2C:FRAMe<n>:STARt?
Returns the start time of the specified frame.
Suffix:
<b>
.
1 | 2
<n> *
Selects the frame.
Return values:
<StartTime> Range: depends on sample rate, record length, and time
base
Increment: depends on the time base
Default unit: s
Usage: Query only
BUS<b>:I2C:FRAMe<n>:STOP?
Returns the end time of the specified frame.