Remote Control Commands
R&S
®
RTB2000
437User Manual 1333.1611.02 ─ 04
Suffix: 
<b>
.
1..2
Selects the parallel bus.
Parameters:
<BusWidth> Maximum number is the number of input channels.
Range:  1  to  4
Increment:  1
*RST:  4
Default unit: Bit
BUS<b>:CPARallel:WIDTh <BusWidth>
Sets the number of lines to be analyzed for the parallel clocked bus.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
Parameters:
<BusWidth> Range:  1  to  15 (clock only) or 14 (clock and CS)
Increment:  1
*RST:  4
Default unit: Bit
BUS<b>:PARallel:DATA<m>:SOURce <DataSource>
BUS<b>:CPARallel:DATA<m>:SOURce <DataSource>
Defines the logic channel that is assigned to the selected bit.
Use the command for each bit of the bus.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
<m> Sets the bit number.
Parameters:
<DataSource> D0..D15
Example: 
BUS:PARallel:Width 4
BUS:PARallel:DATA0:SOURce D8
BUS:PARallel:DATA1:SOURce D9
BUS:PARallel:DATA2:SOURce D10
BUS:PARallel:DATA3:SOURce D11
Mixed Signal Option (Option R&S RTB-B1)