Remote Control Commands
R&S
®
RTB2000
438User Manual 1333.1611.02 ─ 04
15.12.2.2 Parallel Clocked Bus Configuration
BUS<b>:CPARallel:CLOCk:SOURce............................................................................... 438
BUS<b>:CPARallel:CLOCK:SLOPe.................................................................................438
BUS<b>:CPARallel:CS:ENABle.......................................................................................438
BUS<b>:CPARallel:CS:SOURce..................................................................................... 438
BUS<b>:CPARallel:CS:POLarity..................................................................................... 439
BUS<b>:CPARallel:CLOCk:SOURce <ClockSource>
Selects the logic channel that is used as clock line.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ClockSource> D0..D15
*RST:  D0
BUS<b>:CPARallel:CLOCK:SLOPe <ClockSlope>
Selects if the data is sampled on the rising or falling slope of the clock, or on both
edges (EITHer). The clock slope marks the begin of a new bit.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ClockSlope> POSitive | NEGative | EITHer
BUS<b>:CPARallel:CS:ENABle <ChipSelectEnable>
Enables and disables the chip select line.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
Parameters:
<ChipSelectEnable> ON | OFF
*RST:  ON
BUS<b>:CPARallel:CS:SOURce <ChipSelectSource>
Selects the logic channel that is used as chip select line.
Suffix: 
<b>
.
1..2
Selects the parallel bus.
 
Mixed Signal Option (Option R&S RTB-B1)