Remote Control Commands
R&S
®
RTB2000
449User Manual 1333.1611.02 ─ 04
Parameters:
<State> ON | OFF
Range: OFF
PGENerator:PATTern:STIMe <SampleTime>
Sets the time at which each sample is applied for the pattern generator function.
Parameters:
<SampleTime> Numeric value
Range: 2.000E-08 to 4.200E+01
*RST: 2.000E-08
Default unit: s
PGENerator:PATTern:PERiod
<PatternPeriod>
Defines the period of the pattern generator function.
Parameters:
<PatternPeriod> Numeric value (Period = Pattern length * Bit time)
Range: MIN 1 Sample * 20ns = 20ns to MAX 2048 Sam-
ples * 42s = 10416s (approx. 2.89h)
*RST: 2.000E-06
Default unit: s
PGENerator:PATTern:FREQuency <PatternFrequency>
Defines the frequency (period) value of the pattern generator function.
Parameters:
<PatternFrequency> Numeric value
Default unit: Hz
PGENerator:PATTern:ITIMe <IdleTime>
Defines the idle time of the pattern generator function. The idle time can be only
defined with activated BURST function.
Parameters:
<IdleTime> Numeric value
Range: 2.000E-08 to 4.200000000000E+01
*RST: 2.5000000000E-01
Default unit: s
PGENerator:PATTern:BURSt:STATe <BurstState>
Turns the BURST function on or off.
Signal Generation (Option R&S
RTB-B6)