Parameters:
<BurstState> ON | OFF
*RST: OFF
PGENerator:PATTern:BURSt:NCYCle <PatternCycles>
Defines the BURST pattern cycles. The cycles can be only defined with activated
BURST function.
Parameters:
<PatternCycles> Numeric value
Range: 1 to 4096
*RST: 1
PGENerator:PATTern:TRIGger:MODE <TriggerMode>
Defines the arbitrary trigger mode of the pattern generator function.
Parameters:
<TriggerMode> CONTinuous | SINGle
CONTinuous
The CONT function (continuous trigger) issues the pattern con-
tinuously.
SINGle
If the SING setting is activated, the pattern is issued manually.
*RST: CONT
PGENerator:PATTern:TRIGger:SINGle
Manual output of a pattern (single trigger).
Usage:
Event
PGENerator:PATTern:TRIGger:EXTern:SLOPe
Defines the slope of the external arbitrary pattern trigger.
Parameters:
<ExternSlope> POSitive | NEGative | EITHer
POSitive
Rising edge (rise).
NEGative
Falling edge (fall).
EITHer
Rising as well as the falling edge (both).
*RST: POSitive