Protocol Analysis
R&S
®
RTM20xx
204User Manual 1317.4726.02 ─ 01
Fig. 11-16: Decoded LIN signal with frame table and applied label list
gray = synchronization break, synchronization byte, correct checksum
yellow = identifier
green = parity bits
blue = data words (UART words)
Table 11-5: Content of the LIN frame table
Column Description
Start time Time of frame start in relation to the trigger point
ID Identifier value, hexadecimal value
Label Symbolic label, available if a label list was loaded and applied
Length Number of data bytes
Data Hexadecimal values of the data bytes
Chks Checksum value
State Overall state of the frame.
Remote commands:
BUS<b>:LIN:FCOunt? on page 426