Remote Control
R&S
®
RTM20xx
325User Manual 1317.4726.02 ─ 01
REFCurve<m>:VERTical:SCALe <Scale>
Changes the vertical scale of the reference waveform.
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Parameters:
<Scale> *RST: 1
Default unit: V/div
REFCurve<m>:VERTical:POSition <Position>
Changes the vertical position of the reference waveform.
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Parameters:
<Position> *RST: 0
Default unit: div
REFCurve<m>:DATA?
Returns the data of the reference waveform for transmission from the instrument to the
controlling computer. The waveforms data can be used in MATLAB, for example.
To set the export format, use FORMat[:DATA] on page 278.
Suffix:
<m>
.
1..4
Selects the reference waveform, the internal reference storage.
Return values:
<Data> List of values according to the format settings.
Usage: Query only
REFCurve<m>:DATA:HEADer?
Returns information on the reference waveform.
Table 15-14: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval, usually 1. 1