EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Page 347

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
332User Manual 1317.4726.02 ─ 01
Parameters:
<State> ON | OFF
ON
Cursor lines keep their relative position to the waveform.
OFF
Cursor lines remain on their position on the display if the scaling
is changed.
*RST: OFF
CURSor<m>:
RESult?
Returns the measurement result for count, mean, RMS, rise and fall time, peak meas-
urements, and burst width. Make sure to set CURSor<m>:FUNCtion correctly.
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Return values:
<Value> Measurement result
Usage: Query only
CURSor<m>:XDELta:INVerse?
Returns the inverse time difference between the two cursors (1/Δt).
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Return values:
<DeltaInverse> Range: -1,00E+026 to 1,00E+026
Increment: 0.1
*RST: 0
Default unit: 1/s
Usage: Query only
CURSor<m>:XDELta[:VALue]?
Returns the time difference between the two cursors (Δt).
Suffix:
<m>
.
1
The numeric suffix is irrelevant.
Return values:
<Delta> Range: -1,00E+026 to 1,00E+026
Increment: 0.1
*RST: 0
Default unit: s
Command Reference

Table of Contents

Related product manuals