<ReferenceSlope> POSitive | NEGative
Slope of source 2 (second waveform)
*RST:  POS
Firmware/Software:  03.400
MEASurement<m>:RESult[:ACTual]? [<MeasType>]
Returns the result of the specified measurement type.
Suffix: 
<m>
.
1..4
Selects the measurement place.
Query parameters: 
<MeasType> FREQuency | PERiod | PEAK | UPEakvalue | LPEakvalue |
PPCount | NPCount | RECount | FECount | HIGH | LOW |
AMPLitude | MEAN | RMS | RTIMe | FTIMe | PDCYcle | NDCYcle |
PPWidth | NPWidth | CYCMean | CYCRms | STDDev |
CYCStddev | TFRequency | TPERiode | DELay | PHASe |
BWIDth | POVershoot | NOVershoot | TBFRequency | TBPeriod
Specifies the measurement type. See MEASurement<m>:MAIN
on page 336.
Return values: 
<Value> Measurement result. If no measurement was executed, no value
(NAN) is returned.
Usage:  Query only
MEASurement<m>:CATegory?
Returns the measurement category. Currently, the instrument supports only yt-meas-
urements.
Suffix: 
<m>
.
1..4
Selects the measurement.
Return values: 
<Category> AMPTime
AMPtime: yt-measurements
*RST:  AMPT
Usage:  Query only