Remote Control
R&S
®
RTM20xx
355User Manual 1317.4726.02 ─ 01
MASK:STATe <State>
Turns the mask test mode on or off. When turning off, any temporarily stored new masks
are deleted.
Parameters:
<State> ON | OFF
*RST:  OFF
MASK:TEST <Test>
Starts, finishes or interrupts a mask test.
Parameters:
<Test> RUN | STOP | PAUSe
*RST:  STOP
MASK:
LOAD <FileName>
Loads a stored mask from the specified file.
Setting parameters: 
<FileName> String parameter
Path and file name
Usage:  Setting only
MASK:SAVE <FileName>
Saves the current mask in the specified file.
Setting parameters: 
<FileName> String parameter
Path and file name
Usage:  Setting only
MASK:SOURce <Source>
Defines the channel to be compared with the mask.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4
CH3 and CH4 are only available on 4-channel R&S RTM models.
*RST:  CH1
MASK:CHCopy
Creates a mask from the envelope waveform of the test source set with MASK:SOURce.
Command Reference