EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Page 393

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
378User Manual 1317.4726.02 ─ 01
Parameters:
<Polarity> POSitive | NEGative
POSitive = high active
NEGative = low active
*RST: POSitive
BUS<b>:SPI:CLOCk:SOURce <Source>
Selects the input channel of the clock line.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4 | D0..D15
*RST: CH1
BUS<b>:SPI:CLOCk:POLarity <Polarity>
Selects if data is stored with the rising or falling slope of the clock. The slope marks the
begin of a new bit.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Polarity> POSitive | NEGative
POSitive: rising slope
NEGative: falling slope
*RST: NEGative
BUS<b>:SPI:DATA:SOURce <Source>
Selects the input channel of the data line.
Suffix:
<b>
.
1..4
Selects the bus.
Note: SPI/SSPI and UART protocols occupy two bus lines.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4 | D0..D15
*RST: CH1