Remote Control Commands
R&S
®
SMC100A
245Operating Manual 1411.4060.32 ─ 09
:DISPlay:DIALog:CLOSe <DialogId>
Closes the specified dialog. To determine the dialog identifier, use command :
DISPlay:DIALog:ID?.
Setting parameters:
<DialogId> string
Example:
DISP:DIAL:CLOS "<dialog ID>"
Closes the dialog, determined with the "<dialog ID>".
Usage: Setting only
:DISPlay:DIALog:CLOSe:ALL
Closes all open dialogs.
Example:
DISP:DIAL:CLOS:ALL
Usage: Event
:DISPlay:DIALog:ID?
Returns the dialog identifiers of the open dialogs in a string separated by blanks.
Return values:
<DialogIdList> string
Example:
DISP:DIAL:ID?
Response: "<dialog ID(1)> <dialog ID(2)> ...
<dialog ID(n)>"
Returns the dialog identifiers of all opened dialogs.
Usage: Query only
:DISPlay:DIALog:OPEN <DialogId>
Opens the specified dialog. To determine the dialog identifier, use command :
DISPlay:DIALog:ID?.
Setting parameters:
<DialogId> string
Example:
DISP:DIAL:OPEN "<dialog ID>"
Opens the dialog, determined with the "<dialog ID>".
Usage: Setting only
:DISPlay:PSAVe:HOLDoff <HoldoffTimeMin>
Sets the waiting time for the screen-save mode of the display.
DISPlay Subsystem