Loading...
Remote Control Commands
R&S
®
SMC100A
281Operating Manual 1411.4060.32 ─ 09
Example:
MMEM:CDIR '/var/user/ucor'
selects the directory for the user correction files.
CORR:CSET:CAT?
queries which correction tables are available.
Response:UCOR1,UCOR2,UCOR3
the correction tables UCOR1, UCOR2 and UCOR3 are availa-
ble.
Usage: Query only
Manual operation: See "Directory, File List and File Name" on page 112
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency
<Frequency>
Transfers the frequency data to the table selected with :CORRection:CSET:SELect.
The numerical suffix at SOURce must not be used for this command.
Parameters:
<Frequency> Frequency#1[, Frequency#2, ...]
Range: 300 kHz to RFmax (depending on model)
Example:
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:FREQ 100MHz,102MHz,103MHz,...
enters the frequency value in the table ucor1.
Manual operation: See " Edit User Cor. Data - User Correction " on page 141
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
The command queries the number of frequency values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
<Points> integer
Range: 0 to 10000
*RST: 0
Example:
CORR:CSET '/var/user/'
selects the table ucor1.
CORR:CSET:DATA:FREQ:POIN?
queries the number of frequency values in the table ucor1.
Response: 440
the table ucor1 contains 440 frequency values.
Usage: Query only
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Transfers the level data to the table selected with [:SOURce<hw>]:CORRection:
CSET[:SELect].
SOURce Subsystem

Table of Contents